HP-Blood Alcohol
J&W HP-Blood Alcohol GC Columns from Agilent Technologies
Agilent J&W HP-Blood Alcohol is a proprietary phase delivers baseline resolution of ethanol from interferences in biological matrixes such as blood and urine, using appropriate headspace analysis. Every column is use-tested to assure baseline resolution of methanol, ethanol, isopropanol, n-propanol, and n-butanol. Additionally, HP-Blood Alcohol resolves many other common interferents such as acetaldehyde and acetone. An application note shipped with the column describes conditions permitting GC analysis times under four minutes, and total GC cycle-times under 7 minutes, with MDL for ethanol of 0.0003%.
All pictures shown are for illustration purpose only. Actual product may vary.
- Buy
- Features
-
- Reliable blood alcohol analysis
- Excellent confirmation column with DB-ALC2 for method using t-butanol as internal standard
- Each Agilent J&W GC column is tested with the tightest industry QC specifications for column bleed, sensitivity, and efficiency to provide you utmost confidence in qualitative and quantitative results
- For details about the QC test conditions including the test probe mixture and the resulting chromatogram, please refer to the Agilent J&W GC Column Performance Summary shipped with each GC column
Similar Phases: Aucune
- Specifications
-
HP-Blood Alcohol Temperature Specifications
ID (mm) 0.32 mm Length (m) 7.5 m Film (µm) 2.00 µm Temperature Limits (°C) -60 -270/290 °C - Literature
-
-
Agilent GC Column Brochure
Full range of Agilent Technologies GC Columns and Applications
-